IVD Performance Evaluation Under the IVDR
Performance evaluation is the evidence backbone of every IVDR technical file. Under Article 56 and Annex XIII of Regulation (EU) 2017/746, IVD manufacturers must show, with data, that their device does what they claim it does. MedDeviceCorp helps IVD manufacturers plan, generate and document that evidence so it stands up to notified body review.
Get a Performance Evaluation QuoteWhat Is IVD Performance Evaluation?
IVD performance evaluation is a continuous process of assessing and analysing data to establish or verify a device's scientific validity, analytical performance and clinical performance. It is not a one-time report. It starts at design, supports CE marking, and is updated throughout the device's life with post-market data.
The Three Pillars of Performance Evidence
Scientific Validity
Scientific validity shows that the analyte or marker is associated with a clinical condition or physiological state. Evidence typically comes from peer-reviewed literature, consensus guidelines, or proof-of-concept studies.
Analytical Performance
Analytical performance shows that the device correctly detects or measures the analyte. This covers parameters such as analytical sensitivity and specificity, trueness, precision, accuracy, limits of detection and quantitation, measuring range, linearity and interference.
Clinical Performance
Clinical performance shows that the device produces results that correlate with the clinical condition in the intended population and user. Typical parameters include diagnostic sensitivity and specificity, positive and negative predictive values, and expected values in normal and affected populations.
Master Structure of the Analytical Performance Report (APR)
The Analytical Performance Report brings together all analytical performance data for the device and feeds directly into the Performance Evaluation Report. A well-structured APR makes it easy for the notified body to trace every claim back to its evidence. We build APRs around the following master structure:
- Scope & Administrative Identifiers (Device name, UDI-DI, formulation, software version) – defines exactly which device, formulation and software version the report covers.
- Traceability Matrix to the Performance Evaluation Plan (PEP) & GSPR Annex I (9.1a) – links each analytical claim to the PEP and to the general safety and performance requirements.
- Reference Materials, Calibrators & Metrological Traceability – the calibration hierarchy and the reference materials or methods used.
- Analytical Sensitivity: LoB, LoD, and LoQ – limit of blank, limit of detection and limit of quantitation.
- Precision (Repeatability, Reproducibility, Within-Laboratory) – variability across runs, days, sites, lots and operators.
- Trueness, Accuracy, and Method Comparison – agreement with a reference method or comparator device.
- Analytical Specificity (Cross-Reactivity & Exogenous/Endogenous Interference) – the effect of related analytes and interfering substances on results.
- Linearity, Measuring Range, and High-Dose Hook Effect – the reportable range and behaviour at very high analyte levels.
- Specimen Handling, Collection Matrices, and Pre-Analytical Stability – specimen types, collection tubes, transport and storage conditions.
- System Robustness, Carryover, and Reagent Open/Closed-Vial Stability – performance under varied conditions, sample-to-sample carryover and reagent shelf life.
- Statistical Protocols, Deviations, and Notified Body Synthesis – statistical methods and acceptance criteria, study deviations, and a summary conclusion for the notified body.
Studies are typically designed using recognised standards such as the CLSI EP guidelines and ISO 17511 for metrological traceability.
CLSI-Aligned Audit Checklist by Parameter
Notified bodies review analytical performance studies against recognised CLSI and ISO methods. The checklist below shows, for each parameter, the applicable standard, what reviewers check and the non-conformities we see most often.
| Parameter | Applicable Standard | Key Notified Body Review Checkpoints | Common Audit Non-Conformities |
|---|---|---|---|
| Detection Limits (LoB, LoD, LoQ) | CLSI EP17-A2 |
|
|
| Precision | CLSI EP05-A3 |
|
|
| Trueness & Method Comparison | CLSI EP09c CLSI EP29-A |
|
|
| Analytical Specificity & Cross-Reactivity | CLSI EP07-A3 CLSI EP37 |
|
|
| Endogenous & Exogenous Interference | CLSI EP07-A3 CLSI EP37 |
|
|
| Linearity & Measuring Range | CLSI EP06-Ed2 |
|
|
| High-Dose Hook Effect | CLSI EP06 / assay-specific |
|
|
| Matrix Comparison & Anticoagulants | CLSI EP35 |
|
|
| Specimen & Reagent Stability | CLSI EP25-A |
|
|
| Metrological Traceability | ISO 17511:2020 |
|
|
Key IVDR Performance Evaluation Documents
- Performance Evaluation Plan (PEP): defines the intended purpose, claims, evidence needed and methods to obtain it.
- Performance Evaluation Report (PER): summarises scientific validity, analytical and clinical performance data and concludes on conformity.
- Performance study documentation: study plans and reports for analytical and clinical performance studies, including those under Articles 57 to 77 where applicable.
- Post-Market Performance Follow-up (PMPF) plan and report: keep the evidence current after the device is on the market.
- Summary of Safety and Performance (SSP): required for Class C and Class D devices.
How Requirements Change by Risk Class
Every IVD class needs a performance evaluation, but the depth of evidence and level of scrutiny rise with risk. Class D devices may also need to meet common specifications and undergo verification by an EU reference laboratory. For legacy CE-IVD devices, data generated under the old IVD Directive often needs to be supplemented to meet IVDR expectations.
Our IVD Performance Evaluation Services
- Gap assessment of existing performance data against IVDR Annex XIII
- Scientific validity literature reviews
- Performance evaluation plans and reports (PEP and PER)
- Analytical and clinical performance study design and protocol review
- PMPF plans, PMPF reports and SSP preparation
- Support with notified body questions on performance evidence
Performance evaluation is one part of the wider IVDR CE certification process, alongside classification, technical documentation and quality management. Planning it early is the best way to avoid delays during notified body review.
Get Help With Your IVD Performance Evaluation
Tell us about your device, its intended purpose and the data you already have. We will identify the gaps and propose a clear plan for your performance evaluation.
Request an IVDR Performance Evaluation Proposal